DGH DIN-130 Modbus Voltage Input Module

  • DGH

Product Features

  • 8-bit ADC, 8 conversions per second
  • Accuracy (error from all sources between 0-40°C ambient): ±1°C maximum for J, K, T or E and ±2.5°C for R, S, B or C
  • Resolution: 1°C
  • Common mode rejection: 100dB at 50/60Hz
  • Lead resistance effect: <20µV per 350Ω
  • Digital filtering
  • Input impedance of 100MΩ minimum
  • Temperature range of -25°C to +70°C

Applications

  • Non-compliant

The DIN-130 MODBUS thermocouple input modules are part of the DGH’s DIN-100 data acquisition module family. This input module converts analog input signal from the choice of eight different thermocouples (J, K, T, E, R, S, B and C) using the MODBUS RTU protocol to digital data at the rate of 8 conversions per second, storing the latest result in the buffer. The signal is transmitted to the computer or other processor-based controlling equipment using RS-485 port; if this is not available a DIN-191 converter to RS-232 can be used. The DIN-130 module can be directly connected to a sensor and perform signal conditioning, linearisation, scaling and conversion to both ASCII and MODBUS values. For extra security, it utilises a non-volatile EEPROM (Electrically Erasable Programmable Read Only Memory) to store all user-selectable settings, meaning the values are maintained in the event of power loss. The DIN-130 comes pre-set with default values of 300 baud, one start bit, eight data bits, one stop bit, no parity and any address recognized, but these values can be easily changed by the user. The modules require no special software and can be connected and started up quickly and easily, however a utility software is available at no extra charge if required. The DIN-130 can be mounted on a DIN-rail directly, without need for a mounting kit.

  • Single channel analog-to-digital thermocouple input module
  • RS-485 output
  • Choice of 8 different thermocouple inputs: J, K, T, E, R, S, B or C (see specification tab for temperature ranges)
  • Communicates using MODBUS RTU protocol
  • User-selectable features include analog input range, baud rate, parity type and others
  • Accuracy (error from all sources between 0-40°C ambient): ±1°C maximum for J, K, T or E and ±2.5°C for R, S, B or C
  • Resolution: 1°C
  • Lead resistance effect: <20µV per 350Ω
  • Input burnout protection to 250 Vac
  • Input impedance: 100MΩ minimum
  • Open thermocouple indication
  • Overrange indication
  • Automatic cold junction compensation and linearisation
  • 2 programmable single-pole digital filters for smoothing analog data
  • Power requirements: Serial = 0.75W
  • Temperature range -25°C to +70°C

 

The applications for the DIN-130 data acquisition modules include:

  • Process monitoring and control
  • Remote data logging to any host computer
  • Product testing
  • Interface to PLC

 

Specifications

Weight 0.00000 kg
Dimensions 25.40 x 83.80 x 86.40mm
Input/Output

1 Analog Input, 1 Digital Output

Number of Channels

1 Channel

Power Requirements

Unregulated +10V to +30Vdc Serial 0.75W maximum

Series / Model

Thermocouple Type

B (0 to +1820°C), C (0 to +2315°C), E (-100 to +1000°C), J (-200 to +760°C), K (-150 to +1250°C), R (0 to +1750°C), S (0 to +1750°C), T (-200 to +400°C)

DGH DIN-100 Technical Library

We are pleased to provide you with a range of additional content including videos, product datasheets, case studies, white papers and application notes for your reference. Please see below for the latest content available:

 

 

DOCUMENTATION
DGH DIN-100 Series Datasheet
DGH DIN-100 Series User Manual
Command set for DIN-100 Series
MODBUS RTU Function List
MODBUS RTU Protocol Overview

Related Products

  • DGH